EasyManua.ls Logo

Minebea CSD-701B - 9−4−6. Process of communication error

Minebea CSD-701B
157 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
106
946. Process of communication error
The instrument sends back the error command to the host side at the time of occurrence of
communication error or executing error.
2468
@ 0 0 E R 0 1 Terminator
Error
Command No.
ID No.
0031
Error
Command No.
Contents Remarks
01 Error of status of
impossible to execute
During the Function mode, Calibration mode,
Fine adjustment mode and Setting mode.
02 Another error caused from
the instrument
In the case of impossible to execute the
receiving command.
10 Parity error In the case of detecting the parity error.
11 Framing error In case of detecting error for stop bit.
12 Overrun error In the case of reading error for receiving
command.
13 Error of data code, data
length error
The receiving data code and data length are
not the same.
14 No applicable command The receiving command isn’t the same.
In case that the ID number and the Completion code (terminator) is
not detected, the error command will not be returned.
In case that the communication error command is returned from the
instrument, consider its remedy at the host side.

Table of Contents