11. Specifications
11−1. Specifications for analog section
Bridge power supply DC5 V±0.25 V within 60 mA (Changeable to DC2.5 V)
Applicable transducers Up to 4 pieces of strain gage applied transducers(350 Ω) can be
connectable.
Input range F.S. setting is available at the input range from 0.4 mV/V
to ±3.1 mV/V.(When bridge power supply is DC5 V.)
Output range DC+5 V Load resistance is 5 kΩ or more.
F.S. setting is available in Function.
Output times 4 times/s、20 times/s、50 times/s or 100 times/s can be changeable.
(Synchronous with display rate.)
Output resolution Approx.1/10 000
Zero adjustment range −0.3 mV/V to 2.4 mV/V
Non−linearity
Display 0.05 %F.S.
Output 0.05 %F.S.
Temperature coefficient
Zero point ±1 μV/℃(Input conversion, at the time of F.S. setting
at the input from 0.5 mV/V to 3.1 mV/V)
Sensitivity ±0.02 %F.S./℃(Input conversion, at the time of F.S. setting
at the input from 0.5 mV/V to 3.1 mV/V)
Input noise ±1 μVp−p or less
(With the default setting of digital filter and stabilized filter)
Input filter 1 Hz(At the “0” setting of the digital filter and the stabilized filter)
A/D sampling 20 times/s
CHECK Approx.0.4 mV/V
(Setting can be available at the approx. 0.2 mV/V interval within the
range from approx.0.2 mV/V to 1.4 mV/V.)
※The extension cable should be applied to Minebea’s standard cable
CAB-502(4 cores) within the length of 30 m.
※Not applicable when the zener barrier is used.
Analog peak hold Response speed:Corresponds to the characteristics of input filter.
Accuracy:0.1 %F.S. or less
11−2. Specifications for digital section
Load display
Display range −9 999 to 99 999(×2, ×5or×10 is available.)
Display increment 1 (2, 5 or 20 changeable)
Display 7 segment red LED, with 17 mm character’s height
Over display “−OL” displays at the time of minus(−) over, and “OL” displays at
the time of plus (+) over.
Status display ◎, HOLD, CHECK, PEAK, A/Z
Judgement display S1 and S2,
Display rate 4 times/s(20 times/s changeable.)
Decimal point display Non, 10
1
,10
2
,10
3
or 10
4
changeable.