Minebea CSD-701B - 5−2−6. Calibration procedure for re-registration of zero point only

Minebea CSD-701B
157 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
526. Calibration procedure to apply registration again for zero point only
(1) Procedures by key operation
Warning During the execution of calibration, be sure to set the Tare weight
cancellation clear, cancellation of the Compensated data at Zero set
(Execution of F98), and set OFF the Zero tacking (Set the F08 and
F09 to “00000”.) and set the Peak OFF.
During the calibration procedure, press the key to interrupt the
calibration. The calibration data will keep the same condition as it is
entered before, then returns to the Measurement mode.
When the key is pressed with the load display of “FUNC”, the
display will change as the following arrow marks indicate at every
time the key is pressed. However, every time the is pressed, the
display will change as the reverse direction of the following arrow
marks.
“FUNC”“CCAL”ACAL”“LCAL”“ZERO”“SPAN”
“TARE”“CHEK”“MONT” “VCAL”“VADJ”“FUNC”
“CCAL”→・・・・・(Hereinafter, it will repeat.)
Procedures
1
Press the key for about one second.
The load display will show “FUNC”.
2
Press the key six times.
It will make the load display proceeded as
“FUNC”“CCAL”ACAL”“LCAL”
“ZERO”“SPAN”TARE”.
Here, set the initial load application.