6−3−2. When operated in another modes
(1) Increment of set value
By pressing the
key with the condition of displaying various kinds of set values, the set
value will increase per one count from 0, 1, 2, 3, 4, 5, 6, 7, 8, 9 and 0 again in order.
(2) Increment at the time of fine adjustment
When the
key is pressed at the time of making fine adjustment on ZERO, SPAN and
analog output, the targeted value will increase.
● In the operation method of the paragraph 6−3−2(2), when the key is
pressed for more than about 1 second, each operation will be executed
continuously at a constant interval without ON/OFF operation of the
key.
6−4. key
6−4−1. When operated in the Measurement mode.
(1) When the peak function is selected (Lighting off the Status display “◎”).
Changeover the display of peak value/track value.
(2) When the A/Z function is selected (Lighting up the Status display “◎”).
At the same time the Tare weight cancellation (A/Z) works and the load display value become
the net weight display value, the “A/Z” on the status display will light up and load display
becomes “0”.
6−4−2. When operated in another modes
(1) Calibration
The set value is initialized in each procedure of the calibration method.
(2) Function mode
In the function mode, the function number and the set value of the function are compulsorily
made “0”.
(3) Comparator setting
The set value is compulsorily made “0” with the set value of the comparator can set.