117
Displays the load.
Replace fuse by
referring to the
paragraph 131.
Display fluctuates
abnormally.
Load display
doesn’t vary.
Execute the calibration again
according to the procedures in
the paragraph 5.
Inform Minebea about the contents of
failure and situation at site in details.
YES
NO
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
8
7
1
Displays
optional values.
OL display
Same situation
Start measurement.
OL display
7
6
Fuse is broken.