5.6 Command Explanation
5-51 ZM2371/ZM2372
<limit comparison result>
[,<primary parameter comparison result>] [,<secondary parameter
comparison result>]
Only the comparison result of the parameter that enabled the
limit comparison function is output.
0 Limit comparison function is disabled
1 IN (within the range of upper limit and lower limit)
2 HI (larger than upper limit)
4 LO (smaller than lower limit)
Query example: :FETC?
Response example: +0,+3.14159E-06,+1.20000E-02,+2
(When comparator function enabled, C-D measurement, deviation not
displayed, ASCII format)
Measurement status = 0 (no error), C = 3.14159E-06 (3.14159F),
D = 1.20000E-02 (0.01200), Comparator’s bin sorting result = 2 (BIN2)
Remarks: When the measurement status is other than 0 (error), the measured
value of primary parameter and secondary parameter is 9.9E+37, and
the comparator result is 11 (16 when bin extended function is enabled).
When the measurement status is 1 or 3, the limit comparison result is 2
(HI).