5.6 Command Explanation
5-57 ZM2371/ZM2372
[:SENSe]:CORRection:CKIT:STANdard3 <primary parameter>,<secondary parameter>
[:SENSe]:CORRection:CKIT:STANdard3?
Description: Sets/queries standard value for LOAD correction.
Set the format of standard value additionally with
[:SENSe]:CORRection:CKIT:STANdard3:FORMat command.
Parameters: <primary parameter> {numeric value, range 0, ±(1E-16 -
9.99999E+11) }
<Secondary parameter> {numeric value, range 0, ±(1E-16 -
9.99999E+11) }
Actually set value follows the format of standard value and it is rounded
to the range same as set by the panel operation. The unit follows the
format of standard value.
Setting example: :CORR:CKIT:STAN3 12.3456E-06, 0.01234
Sets standard value for LOAD correction to Cs = 12.3456F, D = 0.01234.
(when the format is CSD)
Response: <primary parameter>,<secondary parameter> {numeric value, format
NR3}
Query example: :CORR:CKIT:STAN3?
Response example: +1.23456E-05,+1.23400E-02
Standard value for LOAD correction is Cs = 1.23456E-05 F (12.3456F),
D = 1.23400E-02 (0.01234). (when the format is CSD)
[:SENSe]:CORRection:CKIT:STANdard3:FORMat {CPD|CSD|RCP|RLS|RX|ZPH}
[:SENSe]:CORRection:CKIT:STANdard3:FORMat?
Description: Sets/queries the format of standard value for LOAD correction and
LOAD correction value.
Parameters: CPD Cp, D (in either case, the order is primary parameter and
secondary parameter)
CSD Cs, D
RCP Rp, Cp
RLS Rs, Ls
RX Rs, X
ZPH Z, θ
Setting example: :CORR:CKIT:STAN3:FORM CSD
Sets the format of standard value for LOAD correction and LOAD
correction value to Cs,D.
Response: {CPD|CSD|RCP|RLS|RX|ZPH}
Query example: :CORR:CKIT:STAN3:FORM?
Response example: CSD
The format of standard value for LOAD correction and LOAD correction
value is Cs,D.