5.6 Command Explanation
5-66 ZM2371/ZM2372
[:SENSe]:FUNCtion:CONCurrent {ON|OFF|1|0}
[:SENSe]:FUNCtion:CONCurrent?
Description: Sets/queries how many measurement functions, two or only one, are
specified.
The measurement function are specified with [:SENSe]:FUNCtion[:ON]
command.
Parameters: {ON|1} Specifies two measurement functions.
{OFF|0} Specifies only one measurement function. (Initial value,
*RST)
Setting example: :FUNC:CONC ON
Sets so as to specify two measurement functions.
Response: {1|0}
Query example: :FUNC:CONC?
Response example: 1
A setting is made so as to specify two measurement functions.
[:SENSe]:FUNCtion[:ON] <measurement function>
[:SENSe]:FUNCtion[:ON]?
Description: Sets/queries the measurement functions.
Set in advance the number of measurement functions to be specified
with [:SENSe]:FUNCtion:CONCurrent command.
Parameters: 1) When only one measurement function is specified
<measurement function> <description>
"FIMPedance" Impedance measurement (series
equivalent circuit)
"FADMittance" Admittance measurement (parallel
equivalent circuit)
2) When two measurement functions are specified
<measurement function> <description>
"FIMPedance","FRESistance" Impedance measurement (series
equivalent circuit) and DC
resistance measurement
"FADMittance","FRESistance" Admittance measurement (parallel
equivalent circuit) and DC
resistance measurement
Setting example: :FUNC 'FIMP','FRES'
Specifies two measurement functions of impedance measurement and
DC resistance measurement.
Response: <measurement function>
{"FIMP"|"FADM"|"FIMP","FRES"|"FADM","FRES"}
Query example: :FUNC?
Response example: "FIMP","FRES"
Two measurement functions of impedance measurement and DC
resistance measurement are specified.
Remarks: The FIMPedance / FADMittance setting from the remote control and the
series equivalent circuit / parallel equivalent circuit setting from the
panel are interlocked (same).