5.6 Command Explanation
5-65 ZM2371/ZM2372
[:SENSe]:FRESistance:RANGe:AUTO {ON|OFF|1|0}
[:SENSe]:FRESistance:RANGe:AUTO?
Description: Sets/queries automatic switching of DC resistance measurement range.
Parameters: {ON|1} Enables automatic switching of DC resistance measurement
range.
{OFF|0} Disables automatic switching of DC resistance measurement
range, and fixes the DC resistance measurement range
(HOLD).
Setting example: :FRES:RANG:AUTO ON
Sets so as to enable automatic switching of DC resistance measurement
range.
Response: {1|0}
Query example: :FRES:RANG:AUTO?
Response example: 1
Automatic switching of DC resistance measurement range is enabled.
[:SENSe]:FRESistance:RANGe[:UPPer] <range>
[:SENSe]:FRESistance:RANGe[:UPPer]?
Description: Sets/queries the DC resistance measurement range.
Parameters: <range>
{numeric value, range 100E-3|1|10|100|1E+3|10E+3|100E+3|1E+6,
unit: }.
Parameter specifying method is same as that for the measurement
range of AC impedance.
For datails [:SENSe][:FIMPedance]:RANGe[:UPPer]
Setting example: :FRES:RANG 100E-1
Sets the measurement range to 100m.
Response: {numeric value, format NR3}
Query example: :FRES:RANG?
Response example: +1.00000E-01
The DC resistance measurement range is 1.00000E-01 (100m).
Remarks: Automatic switching of DC resistance measurement range is
automatically disabled if specific DC resistance measurement range is
set. The DC resistance measurement range cannot be specified or fixed
from the panel.