5.6 Command Explanation
5-68 ZM2371/ZM2372
:SOURce:FREQuency[:CW] <frequency>
:SOURce:FREQuency[:CW]?
Description: Sets/queries the measurement frequency.
Parameters: <frequency>
{numeric value, range 0.001 to 100.00E3, resolution 5 digits (<1mHz
when 10Hz), unit : Hz} Suffix M (10
-3
) / K (10
3
), and unit: Hz ,
MAX/MIN can be used.
Example: 0.12K (= 120), 1KHZ (= 1E3)
Setting example: :SOUR:FREQ 1000
Sets the measurement frequency to 1kHz
Response: <frequency> {numeric value, format NR3}
Query example: :SOUR:FREQ?
Response example: +1.00000E+03
The measurement frequency is 100.00E+03 Hz (1kHz).
:SOURce:RESistance[:LIMit]:LOW <resistance>
:SOURce:RESistance[:LIMit]:LOW?
Description: Sets/queries the minimum output impedance of drive signal source.
Parameters: <resistance> { numeric value, range 5, 25, 100, unit: }
As a unit, OHM can be used. Parameter example: 5OHM (= 5).
If an arbitrary value is given, it becomes a nearest value.
Setting example: :SOUR:REG:LOW 25
Sets the minimum output impedance to 25.
Response: <resistance> {numeric value, format NR3}
Query example: :SOUR:REG:LOW?
Response example: +2.50000E+01
The minimum output impedance is 25.