EasyManua.ls Logo

NF ZM2371 - Page 196

NF ZM2371
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.6 Command Explanation
5-69 ZM2371/ZM2372
:SOURce:VOLTage:ALC[:STATe] {ON|OFF|1|0}
:SOURce:VOLTage:ALC[:STATe]?
Description: Sets/queries whether constant voltage drive is executed or not.
Parameters: {ON|1} Enables the constant voltage function. (Control so that
monitored value falls within 1% of set value)
{OFF|0} Disables the constant voltage function.
Setting example: :SOUR:VOLT:ALC OFF
Sets so as to disable the constant voltage function.
Response: {1|0}
Query example: :SOUR:VOLT:ALC?
Response example: 0
The constant voltage function is disabled.
Remarks: The range in which ALC (automatic level control) functions actually is
limited.
For details 3.5.6.3 Measurement Range
The constant current function is automatically disabled if the constant
voltage function is enabled.
:SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude] <voltage>
:SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Description: Sets/queries the measurement voltage level.
Parameters: <voltage> {numeric value, range 0.010 to 5.00, setting resolution 3
digits (0.001 when <1), unit : Vrms} Suffix M (10
-3
), and unit V ,
MAX/MIN can be used.
Example : 1000M (= 1), 500MV (= 0.5)
Setting example: :SOUR:VOLT 0.5
Sets the measurement voltage level to 0.5 Vrms.
Response: <voltage> {numeric value, format NR3}
Query example: :SOUR:VOLT?
Response example: +5.00000E-01
The measurement voltage level is 5.0000E-01 Vrms. (0.5 Vrms.).

Table of Contents

Related product manuals