8.7 Performance Testing
8-10 ZM2371/ZM2372
Cable length 1m, Measurement frequency 100kHz, Measurement signal level 1 Vrms
Calibration value
of standard “A”
For 1kΩ, 10kΩ, or 100kΩ range, use the standard resistor calibrated by 100kHz, if
available. Otherwise, use the standard capacitor. Use either one.
Cable length 2m, Measurement frequency 10kHz, Measurement signal level 1 Vrms
Calibration value
of standard “A”
Cable length 4m, Measurement frequency 1kHz, Measurement signal level 1 Vrms
Calibration value
of standard “A”
Measurement signal level 0.1 Vrms, Measurement frequency 1kHz, Cable length 0m
Calibration value
of standard “A”
Measurement signal level 5 Vrms, Measurement frequency 1kHz, Cable length 0m
Calibration value
of standard “A”