Alarms, Notications, and Other Messages 9
100
9 Alarms, Notications, and Other Messages
PDM message Beeps Description What to do
Reminder:
No active Pod.
* 2 sets of 4
beeps, repeat
every 15 min.
PDM beeps.
You need to activate a
new Pod to begin basal
insulin delivery.
Press OK.
Activate a new
Pod.
Reminder:
Always check BG
after a pod change.
Check infusion site
and ensure cannula
is properly inserted.
* 2 sets of 3
beeps, repeat
every 5 min.
PDM beeps.
Reminds you to check
your blood glucose 90
minutes aer a Pod
change.
Press OK.
Check blood
glucose.
Reminder:
Check BG.
*
§
2 sets of 3
beeps, repeat
every 5 min.
PDM beeps.
Reminds you to check
your blood glucose
aer a bolus.
Press OK.
Check blood
glucose.
Reminder:
Missed bolus.
Meal bolus not
delivered during
reminder period
<start time> - <end
time>
*
§
2 sets of 3
beeps, repeat
every 15 min.
PDM beeps.
You have not delivered
a bolus within the
time period you have
specied.
Press OK to
acknowledge the
bolus reminder.
Low PDM batteries.
Change batteries
soon.
No beeps. e PDM batteries are
low.
Replace the
batteries as soon
as you can.
<User-dened
custom message>
*
§
2 sets of 3
beeps, repeat
every 15 min.
PDM beeps.
Custom reminder
with a message of your
choice.
Press OK.
* To set to vibrate and/or beep, see "Vibrate and sound options" on page 62.
§ To turn on or o, see "BG reminders" on page 77, "Bolus reminders" on page 78,
"Custom reminders" on page 80.