71
Adjusting Settings 77 Adjusting Settings
You can set dierent Target BG and Correct Above values for dierent times of
day. To edit or add Target BG or Correct Above segments:
1. Home > Settings > System setup > Bolus/basal/calcs
> Ratios/factors/targets > Target BG
2. Follow the on-screen instructions, or, for detailed instructions see "Adding
time segments" on page 19.
3. Press Done, then Save when nished modifying the time segments.
Minimum BG for Calcs
Minimum BG for Calcs is a threshold that you set. If your BG reading is below
your Minimum BG for Calcs, the Bolus Calculator is disabled and does not
calculate a bolus. is value can range from 2.8 - 3.9 mmol/L and is a xed value
throughout the day.
1. Home > Settings > System setup > Bolus/basal/calcs
> Ratios/factors/targets > Min BG for calcs
2. Use the Up/Down button to specify the minimum blood glucose for use by
the Bolus Calculator, then press Next.
IC Ratio
e IC Ratio (Insulin to carbohydrate ratio) denes how many grams of
carbohydrates are handled by one unit of insulin. e Bolus Calculator uses your
IC Ratio to calculate a meal bolus when you are going to eat. Your IC Ratio can
range from 1 to 150.
You can set dierent IC Ratios for dierent times of day, up to a total of eight. To
edit existing time segments or add segments:
1. Home > Settings > System setup > Bolus/basal/calcs
> Ratios/factors/targets > IC ratio
2. Follow the on-screen instructions, or, for detailed instructions see "Adding
time segments" on page 19.
3. Press Done, then Save when nished modifying the time segments.