Appendix
Summary of Settings and Options
e options for the various Omnipod® Insulin Management System settings are:
Time 12-hour or 24-hour clock
Date MM/DD/YY
DD/MM/YY
MM.DD.YY
DD.MM.YY
YY-MM-DD
Maximum Basal Rate 0.05-30 U/hr. Default is 3.00 U/hr
Basal rate U/hr. Range: 0.05 U/hr to Maximum Basal Rate in 0.05
U/hr increments.
Basal Programs 7
Basal rate segments 24 per program
Temp basal %, U/hr, or O. Default is O
Duration: 30 min to 12 hrs in 30-min increments
Temp basal (set to %) Range: 0 U/hr ("O") to 95%MORE of current basal
rate in 5% increments. Cannot exceed Maximum Basal
Rate.
Temp basal (set to U/hr) Range: 0 0 U/hr ("O") to Maximum Basal Rate
Temp basal presets 7
BG Goal Range for
blood glucose history
Lower and upper limits: 3.9 to 11.1 mmol/L in 0.1
mmol/L increments
BG reminder On or O. Default is O.
Maximum of 4 active at one time
Reminder can occur between 30 min to 4 hrs aer
bolus is started. Set in 15-minute increments.
Custom reminder Maximum of 4. Set to Daily, One time only, O
BG Meter sound On or O. Default is On.
Bolus Calculator On or O. Default is On.
Target BG value 8 time segments; 3.9 to 11.1 mmol/L in 0.1 mmol/L
increments
Correct Above threshold 8 time segments; Target BG to 11.1 mmol/L in 0.1
mmol/L increments
Minimum BG for Calcs 2.8 to 3.9 mmol/L in 0.1 mmol/L increments
Default is 3.9 mmol/L