EasyManua.ls Logo

OPTIMedical OPTI CCA-TS - Page 125

OPTIMedical OPTI CCA-TS
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS 7-7
7 DIAGNOSTICS AND TROUBLESHOOTING
The optics or cassette are dirty.
Remove the cassette. Inspect the cassette and
optics on bottom and top plate.
Clean, if necessary.
• Reinsert the cassette and press
to
rerun the test.
The cassette has been holding the calibration for
more than 10 minutes without a sample being
attached.
• Press
and discard the cassette.
A sample error has occurred. This may be due to a
clot or large air bubble if two or more sensors are
unstable.
• Press
and discard the cassette.
• Check the sample and rerun with a new cassette.
The system was not able to aspirate enough
contiguous sample uid to cover the optode sensors
after multiple aspiration attempts. If a bubble
was detected, the system attempted to restart the
aspiration and was not able to aspirate enough
sample.
• Press
, and discard the cassette.
The bar code was invalid. The OPTI CCA-TS either
misread the bar code label or it is an invalid bar code
for the OPTI CCA-TS.
• Press
to retry.
• If the error message appears again, check the
product package for intended use.
• Check the bar code scanner (see Section 7.2.11).
• Clean the bar code scanner. Using a lint-
free cloth dampened with a dilute alcohol or
ammonia-based cleaner, gently wipe the face of
the scanner clean.
• Retry the bar code.
OK
ERROR! - Dirty Optics.
Clean Cassette/Optics.
OK
ERROR! - Dirty Optics.
Clean Cassette/Optics.
OK
ERROR! - Calibration Expired.
Discard Cassette.
OK
ERROR! - Calibration Expired.
Discard Cassette.
ERROR! - Unstable Sensors.
Discard Cassette.
OK
ERROR! - Unstable Sensors.
Discard Cassette.
OK
ERROR! - Short Sample.
Discard Cassette.
OK
ERROR! - Short Sample.
Discard Cassette.
OK
ERROR! - Invalid Barcode.
OK
ERROR! - Invalid Barcode.
OK

Table of Contents