Operator’s Manual – OPTI CCA-TS 7-9
7 DIAGNOSTICS AND TROUBLESHOOTING
The SRC expiration date has been reached.
• Press and remove the SRC.
• If the error message appears again, check the
date in <System ->Time and Date>.
• Congure a new SRC under <Setup> and retry.
A measurement of QC materials, either liquid or
SRCs, was attempted prior to setting up.
• Press
to continue.
• Congure the SRCs and/or liquid QC material
under <Setup> and retry.
The analyzer is unable to calibrate the sample light
gates due to dirty optics or cassette.
• Remove and discard the cassette. Inspect and
clean the optics glass and inside the sample
measurement chamber top cover.
• Press
to continue.
• Check the LEDs (See Section 7.2.4).
The Operator ID already exists in the database.
• Press
to continue.
• Enter a unique Operator ID.
The PIN number does not exist in current Secure Op.
ID database.
• Press
to continue.
• Retry with a valid PIN number.
The PIN number already exists in database.
• Press
to continue.
• Enter a unique PIN number.
ERROR! - Expired SRC.
Check Time and Date.
OK
ERROR! - Expired SRC.
Check Time and Date.
OK
ERROR! - Need Setup-QC
OK
ERROR! - Need Setup-QC
OK
ERROR! - Dirty Optics.
Discard Cassette.
OK
ERROR! - Dirty Optics.
Discard Cassette.
OK
ERROR! - Op ID already Exists.
OK
ERROR! - Op ID already Exists.
OK
ERROR! - Invalid PIN.
OK
ERROR! - Invalid PIN.
OK
ERROR! - PIN already Exists.
OK
ERROR! - PIN already Exists.
OK