Operator’s Manual – OPTI CCA-TS 7-11
7 DIAGNOSTICS AND TROUBLESHOOTING
The gas cylinder is empty.
• Replace the gas cylinder and press .
If SRC QC lockout has been activated in <Setup>,
this message will be displayed if SRCs have not
been run within the specied time.
• Press
and run SRCs.
The instrument has detected an internal error.
• Discard the cassette.
• Turn the power off, wait 30 seconds and then
turn the power back on.
Patient, QC and other databases were deleted.
• Press
and the instrument will
reinitialize.
This error message will appear if you try to run
lactate cassettes if the lactate parameter has not been
set up on your analyzer.
• Refer to section 7.2.20 for instructions to set up
the lactate parameter.
If QC lockout has been activated in <Setup> this
message will be displayed if controls have not been
run within the specied time.
• Press
and run control materials.
The temperature is out of range during any kind of
measurement.
• Press
and continue.
• If the error message appears again, check the
temperature under <System - Diagnostics>.
STOP! - Low Gas.
Replace Now.
OK
STOP! - Low Gas.
Replace Now.
OK
STOP! - Need New SRC Data.
Must Run SRCs.
OK
STOP! - Need New SRC Data.
Must Run SRCs.
OK
STOP! - System Error.
Cycle Power for Reset.
STOP! - System Error.
Cycle Power for Reset.
STOP! - Memory Error.
Database Deleted.
OK
STOP! - Memory Error.
Database Deleted.
OK
STOP! - Lactate Setup Required.
Lactate Setup . . .
OK
STOP! - Lactate Setup Required.
Lactate Setup . . .
OK
STOP! - Need New QC Data.
Must Run Controls.
OK
STOP! - Need New QC Data.
Must Run Controls.
OK
STOP! - Temperature Out of Range.
OK
STOP! - Temperature Out of Range.
OK