VR2 CONTROL SYSTEM
SK77898/2
PG DRIVES TECHNOLOGY
118
If the trip is still present after the above checks have been made, then the Power
Module may be defective. Refer to Section 5.
3.16 Current Limit Active
This occurs when the control system operates above the Current Limit Threshold for a
period of time greater than the Current Limit Time.
This has been designed to notify the Healthcare Technician that the control system
has opperated outside of its programmed range.
3.17 High Temperature
This occurs when the control system reaches its Temperature Threshold and thus
becomes to hot. The controller goes out of drive into standby to allow the controller to
cool down.
An entry is made in the system log each time the controller gets too hot and goes out
of drive.