Transducer Care and Maintenance
HD11 XE Getting Started
4535 612 62651
8
162
General Cleaning Procedures for All Transducers
These general cleaning instructions are indicated for all transducers. It is impor-
tant that you clean the transducer and cable according to the following proce-
dures.
CAUTIONS
• Transducers must be cleaned after each use. Cleaning the transducer is an
essential step before effective disinfection or sterilization. Be sure to follow
the manufacturer’s instructions when using disinfectants.
• Do not allow sharp objects, such as scalpels or cauterizing knives, to touch
transducers or cables.
• When handling a transducer, do not bump the transducer on hard surfaces.
• Do not use a surgeon’s brush when cleaning transducers. The use of even soft
brushes can damage transducers.
• During cleaning, disinfection, and sterilization, orient the parts of the trans-
ducer that must remain dry higher than the wet parts until all parts are dry.
This will help keep liquid from entering permeable areas of the transducer.
➤ To clean a transducer
1. After every patient exam, wipe the ultrasound transmission gel off of the
transducer.
2. Disconnect the transducer from the system and remove any transducer cov-
ers, biopsy guides, or tip guards.
3. Use a soft cloth lightly dampened in a mild soap or an enzymatic cleaner (in
accordance with the manufacturer’s instructions) to remove any particulate
matter or body fluids that remain on the transducer or cable. Enzymatic
cleaners should have a pH of 6.0 to 8.0. These cleaners are further diluted
during use. For a list of approved enzymatic cleaners, see Table 8-2.
4. To remove remaining particulate and cleaning residue, rinse thoroughly with
water up to the immersion point shown in Figure 8-3.
5. Wipe with a dry cloth.