Transducer Care and Maintenance
Disinfecting Transducers with Wipes and Sprays (Low to
Intermediate-Level Disinfection)
To disinfect transducers, you can use either an immersion method or a wipe
method with a disinfectant recommended by Philips Ultrasound. Use the method
that is biologically appropriate, as described in Table 8-1.
This section provides instructions on using the wipe or spray method. For infor-
mation on the immersion method, see “Disinfecting Transducers by Immersion
(High-Level Disinfection)” on page 167.
➤ To u s e t h e w i p e o r s p r a y m e t h od to disinfect transducers
NOTE
Transducers can be disinfected using the wipe method only if the product labeling
of the compatible disinfectant you are using indicates it can be used with a wipe
method.
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting any
equipment.
1. After cleaning the transducer and cable, wipe or spray the transducer and
cable with a low-level disinfectant. Allow for the manufacturer’s recom-
mended contact time. For a list of compatible disinfectants, see Table 8-2.
2. Remove any residue with a water-moistened soft cloth. Do not allow any
solutions to air dry on the transducer.
CAUTION
The use of 70% isopropyl alcohol (rubbing alcohol) on transducers is restricted.
Wipe only the distal tip of the transducer up to 2.5 cm (1 in) from the strain
relief/housing joint with an isopropyl alcohol solution. Do not wipe the strain
relief/housing joint, the strain relief, or the cable with isopropyl alcohol. Isopropyl
alcohol can cause damage to these parts of the transducer. This damage is not
covered by the warranty or your service contract. See Figure 8-1.