Transducer Care and Maintenance
HD11 XE Getting Started
4535 612 62651
8
168
CAUTIONS
• Using non-recommended disinfectants, incorrect solution strengths, or
immersing a transducer deeper or for a longer contact time can damage or
discolor the transducer and will void the transducer warranty.
• Do not immerse transducers longer than the minimum time needed for your
level of disinfection. For information on the levels of disinfection require-
ments, see Table 8-1.
➤ To disinfect transducers by using the immersion method
1. Clean the transducer according to the “General Cleaning Procedures for All
Transducers” on page 162.
2. Mix the disinfection solution compatible with your transducer (see Table 8-2)
according to label instructions for solution strength. A disinfectant listed in
Table 8-2 with the footnote “FDA 510(k) cleared” is recommended in the
U.S.
1
3. Immerse the transducer into the appropriate disinfectant for your transducer
as shown in Figure 8-3.
Follow the instructions on the disinfectant label for the duration of transducer
immersion. Do not immerse transducers longer than the minimum time
needed for your level of disinfection.
4. Using the instructions on the disinfectant label, rinse the transducer up to the
point of immersion, then air dry or towel dry with a sterile cloth.
5. Examine the transducer for damage, such as cracks, splitting, fluid leaks, or
sharp edges or projections. If damage is evident, discontinue use of the trans-
ducer and contact your Philips Ultrasound representative.
1. See the list of FDA-cleared sterilants and high-level disinfectants with general claims for pro-
cessing reusable medical and dental devices. For more information, see www.fda.gov/cdrh/
ode/germlab.html