Transducer Care and Maintenance
HD11 XE Getting Started
4535 612 62651
8
174
➤ To sterilize a transducer
1. Clean the transducer according to the “General Cleaning Procedures for All
Transducers” on page 162.
2. Mix the sterilization solution compatible with your transducer (see Table 8-2)
according to label instructions for solution strength. A disinfectant listed in
Table 8-2 with the footnote “FDA 510(k) cleared” is recommended in the U.S.
3. Immerse the transducer in the sterilization solution as shown in Figure 8-4.
4. Follow the instructions on the sterilization label for the duration of trans-
ducer immersion required for sterilization.
5. Remove the transducer from the sterilization solution after the recommended
sterilization time has elapsed.
6. Using the instructions on the sterilization label, rinse the transducer in sterile
water up to the point of immersion, and then air dry or towel dry with a ster-
ile cloth.
7. Examine the transducer for damage, such as cracks, splitting, fluid leaks, or
sharp edges or projections. If damage is evident, discontinue use of the trans-
ducer, and contact your Philips Ultrasound representative.