EasyManuals Logo

Philips Respironics SimplyGo Mini User Manual

Philips Respironics SimplyGo Mini
193 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
31
SimplyGo Mini User Manual
English
3� You can view the type of alarm that occurred and the start time and
date of the alarm on the Alarm Log screen�
4� Press the < and > buttons to navigate through the list of alarms in the
log history� The log will list the 16 most recent alarms
Notes:
The alarm log is maintained between device Power Off and Power On
cycles (the time of Power Off is not logged)�
The most recent alarm events in the log from the current Power On
session may be lost if the device experiences a total loss of power and
improperly shuts down�
The alarm log maintains the most recent 16 alarms and will discard the
oldest data once its memory is full and a new alarm occurs� If less than 16
alarms have occurred, N/A will appear on some of the alarm screens�
Change the Language Selection Setting
To select a new language on the SimplyGo Mini
device, complete the following steps:
1� From the Home screen, click on the Tools icon to
access the Device Tools menu�
2� Press the page left or page right (< >) buttons
to scroll through the menus until the language
selection menu appears, shown at right�
3� Press the Language Selection icon, and then press the up and down
arrows (∧ ∨) to scroll through the list until the desired language appears,
as shown below

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing

Related product manuals