56
SimplyGo Mini User Manual
Expected Service Life
of Battery
300 charge-discharge cycles
Shelf Life of Battery 6 months
AC Power
Type Philips Respironics REF 1116818
Input 100 to 240 VAC, 50/60 Hz
Output 19 VDC, 6�3 A max
Power Consumption
120 W while charging
DC Power
Type
Philips Respironics REF 1116819
Input 12-16 VDC
Output
19 +/- 5% VDC, 6�3 A
Standards Compliance
The device is designed to conform to the following standards:
- IEC 60601-1 Medical Electrical Equipment, Part 1: General requirements
for basic safety and essential performance
-- IEC 60601-1-2 Medical Electrical Equipment, Part 1-2: General Require-
ment for Basic Safety and Essential Performance - Collateral Standard:
Electromagnetic disturbances - Requirements and Tests
- IEC 60601-1-6 Medical electrical equipment - Part 1-6: General Require-
ments for Basic Safety and Essential Performance - Collateral Standard:
Usability
- IEC 60601-1-8 Medical electrical equipment – Part 1-8: General Require-
ments for Basic Safety and Essential Performance – Collateral Standard:
General requirements, tests, and guidance for alarm systems in medical
electrical equipment and medical electrical systems
- IEC 60601-1-11 Medical electrical equipment – Part 1-11: General Require-
ments for Basic Safety and Essential Performance – Collateral Standard:
Requirements for medical electrical equipment and medical electrical

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing