34
SimplyGo Mini User Manual
Technical Alarms
Visual, Audio
Indicators
Description What To Do
1 beep
that repeats every
16 seconds
No Breath Alarm
This alarm occurs when
a breath is not detected
for a period of 2 minutes
or more� Press the yellow
Alarm silence button on
the screen to temporarily
silence the alarm, and
the 2nd screen shown at
left appears� Or, press the
Home button on the device
to return to the Home
screen, and the 3rd screen
will appear with the No
Breath (NB) alarm symbol
in the upper left corner to
indicate the active alarm
state�
This alarm resets itself
as soon as a breath is
detected� If the device is
on battery power and no
breath is detected after
approximately 15 minutes,
the device shuts down
to conserve power� (This
alarm is inactive for ten
minutes after startup�)
Note: This alarm is inactive
for 10 minutes after startup�
Upon startup, this alarm
could take a maximum
of 12 minutes to activate
from the time the alarm
condition is present�
Otherwise, it can take up to
2 minutes to activate�
Check the
connection from
the cannula to the
device�
Make sure the nasal
cannula is properly
positioned on your
face and that you are
breathing through
your nose�
Make sure the
cannula tubing is not
kinked or obstructed�

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing