53
SimplyGo Mini User Manual
English
Oxygen
Concentration
90% +6% - 3% at all settings
Flow Settings and
Pulse Volumes
Per ISO 80601-2-67 test method
+/-15% at STPD*
+/-25% over the rated environmental range
*STPD is 101�3 kPa at an operating temperature of 20 °C, dry
Average of 30 consecutive pulses with 1�5 m length tubing
Specications
Operating
Conditions
Operational temperature:
41° F to 95° F (5° C to 35° C)
Relative humidity: 15% to 93%
Atmospheric Pressure: 70 kPa to 101 kPa
Altitude: 0 to 10,000 ft� (3048 m�)
Storage and Transport
Conditions
-4° F to 140° F (-20° C to 60° C) - unit only
Relative humidity: up to 93%, non condensing
Expected Service Life
of the device and
accessories
5 years
Note: Per ISO 80601-2-69, the Sound Pressure level is 58 +/- 3 dBA and the maximum
Sound Power Level is 66 +/- 3 dBA at a continuous setting of 2 LPM� The maximum
Sound Pressure Level is 51 +/- 2 dBA and the maximum Sound Power Level is 59 +/- 2
at Nominal Pulse Mode Setting of 5 and 20 BPM�