46
SimplyGo Mini User Manual
Caring for your SimplyGo Mini System
Cleaning and Disinfecting the Device
Note: This product is suitable for use with other patients once the
device is cleaned and disinfected according to the procedures
in the “Cleaning, “Disinfection, and Accessories Cleaning and
Maintenance sections of this manual�
Note: The SimplyGo Mini oxygen concentrator does not contain any user-
serviceable parts� All service should be referred to an authorized
service center
Warning: To avoid electrical shock, do not remove the SimplyGo Mini covers�
The covers should only be removed by authorized service personnel�
Do not apply liquid directly to the covers� Do not use alcohol,
solvents, polishes, or any oily substances on the device, as they are
ammable�
Caution: Do not allow liquids into any of the controls, the interior of the
case, or the oxygen tubing connector� If this occurs, contact your
equipment provider for assistance�
Cleaning
The exterior covers of the device should be cleaned weekly and between
patient use and as needed by performing the following steps:
1� Turn the device o and disconnect from the power source before
cleaning�
2� Wipe the device exterior with a cloth slightly dampened with water
and a mild dishwashing detergent� Wipe dry
Storage at temperatures above 113° F (45° C), such as in a hot car, may
degrade battery performance and reduce battery life� Storage at low
temperatures may aect initial battery performance�
The optimal temperature range for charging the batteries is 0° F to 113° F (0° C
to 45° C) at up to 80% relative humidityThe device will not charge the battery
if the battery cells exceed 113° F (45° C)� Battery charging will be inhibited
until the battery temperature returns to 111° F (44° C) or below�

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing