17
SimplyGo Mini User Manual
English
Device Warm-Up
When you turn on your SimplyGo Mini device, it will sense if you are breathing
from it� If you are not yet breathing through the cannula, SimplyGo Mini will
begin to pulse automatically about once every ve seconds� As soon as you
begin breathing through the cannula, the device will begin delivering pulses
based on your breathing� It may take up to 20 minutes for the SimplyGo Mini
output to stabilize within its O
2
purity specications� However, you may begin
breathing from the SimplyGo Mini at any time� During the initial stabilization
period, the Warm-Up symbol is displayed under the setting number
Note: If the device is stored at the minimum storage temperature between
uses, please allow 2�5 hours for the unit to adequately warm up
Note: If the device is stored at the maximum storage temperature between
uses, please allow 2�5 hours for the unit to adequately cool down�
Note: The Low Oxygen Concentration Alarm is inactive for approximately 20
minutes after startup� The No Breath Alarm is inactive for the rst ten
minutes after startup�
Setting up the Cannula and Tubing
The alarm system should be veried before use by the equipment provider
and between users by service personnel in accordance with the SimplyGo
Mini service manual�
However, the user may verify the functionality of the alarm system as follows:
1� Induce a No Flow Alarm condition by turning the device ON and
completely blocking the patient cannula connector of the device�
2� Wait approximately 90 seconds or until the No Flow Alarm occurs�
3� Verify that the No Flow Alarm indicators and screen symbols are
present�
4� Unblock the patient cannula connector for a minimum of 60 seconds�
5� Verify that the No Flow Alarm ceases�

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing

Related product manuals