SimplyGo Mini User Manual
English
Disinfection
The exterior covers of the device should be disinfected weekly and between
patient use as follows:
1� Clean the device as indicated above�
2� To disinfect, use a household chlorine bleach containing 8�25%
sodium hypochlorite� Combine 9 parts water to 1 part bleach�
3� Using a damp cloth with the bleach solution, wipe the exterior
surfaces�
4� Allow the surface to remain damp for 2 minutesWipe dry as
necessary�
Accessories Cleaning and Maintenance
C
arrying Case
The carrying case should be cleaned monthly by performing the following
steps:
1� Remove the SimplyGo Mini device from the carrying case�
2� Use a dry, soft-bristled brush gently to remove any visible soil on the
carrying case before wiping�
3� Use a cloth dampened (not dripping) with a mild, household liquid
dishwashing solution to clean the exterior of the carrying case�
Ensure you remove all visible soil; a damp, soft-bristled brush may be
used if necessary�
4� Perform a ‘rinse’ using a cloth dampened (not dripping) with warm
water to wipe the exterior of the carrying case�
5� Allow the case to air dry
The carrying case should be single use and replaced between patients�
Cannula and Tubing
Clean and replace the cannula and tubing as instructed by the manufacturer
and your equipment provider
Important: Always replace the cannula between patient use
Device Storage
Store your device in a place where it will remain clean and dry

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing