Page #61 background imageLoading...
Page #61 background image
58
SimplyGo Mini User Manual
EMC Information
Your unit has been designed to meet EMC standards throughout its Service Life
without additional maintenanceThere is always an opportunity to relocate your
portable SimplyGo Mini within an environment that contains other devices with
their own unknown EMC behavior� If you believe your unit is aected by locating
it closer to another device, simply separate the devices to remove the condition�
Portable Oxygen Concentrator
The SimplyGo Mini is designed to provide enriched oxygen� If there is a condition
where the device is not providing enriched oxygen, the device will alarm�
Oxygen Conserving
The SimplyGo Mini is designed to provide oxygen in response to the patient’s
breathingThe device includes a sensing circuit that will detect when the patient
has started to inhale� If the SimplyGo Mini is either randomly or not providing
pulse dosed oxygen, this is a sign that the detection sensor is currently having
trouble detecting the patient’s breathing pattern�
If there are errors or alarms aecting the purity or conserving features of the
device, the user should attempt to move the SimplyGo Mini to a dierent area to
determine if the issue is due to electromagnetic interference with other equip-
ment in the vicinity
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
SimplyGo Mini is intended for use in the electromagnetic environment specied below� The user of SimplyGo Mini should
assure that it is used in such an environment
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Radiated Emissions
CISPR 11
Group 1
Class B
The Device uses RF energy only for its internal
function� Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment�
RF Conducted Emissions
CISPR 11
Group 1
Class B
The Device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network�
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Complies
RF Emissions
CISPR25
Class 3 The Device is suitable for use in vehicles and
boats�
Emission of Radio Frequency Energy
RTCA/DO-160G Section 21
Category M This Device is suitable for use onboard
commercial airplanes inside passenger cabin�

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing

Related product manuals