59
SimplyGo Mini User Manual
English
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact
discharges
±2 kV, ±4 kV, ±8
kV, and ±15 kV air
discharges
±8 kV contact
discharges
±2 kV, ±4 kV, ±8
kV, and ±15 kV air
discharges
Floors should be
wood, concrete or
ceramic tile� If oors
are covered with
synthetic material,
the relative humidity
should be at least
35%�
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV at 100 kHz
repetition rate for
power supply lines
±1 kV for input-
output lines; both at
100 kHz repetition
rate
±2 kV at 100 kHz
repetition rate for
power supply lines
N/A - Device does not
have user I/O lines
that are longer than
3m in length�
Mains power quality
should be that of
a typical home or
hospital environment�
Surge
IEC 61000-4-5
±1 kV dierential
mode
±2 kV common mode
±1 kV dierential
mode
N/A Device is Class 2
and does not have an
earth connection�
Mains power quality
should be that of
a typical home or
hospital environment�
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0�5 cycle at 45
degree increments
<5% U
T
(>95% dip in
U
T
) for 1 cycle
70% U
T
(30% dip in
U
T
) for 0�5 seconds
<5% U
T
(>95% dip in
U
T
) for 5 seconds
<5% U
T
(>95% dip in
U
T
) for 0�5 cycle at 45
degree increments
<5% U
T
(>95% dip in
U
T
) for 1 cycle
70% U
T
(30% dip in
U
T
) for 0�5 seconds
<5% U
T
(>95% dip in
U
T
) for 5 seconds
Mains power quality
should be that of
a typical home or
hospital environment�
If the user of the
device requires
continued operation
during power mains
interruptions, it
is recommended
that the device be
powered from an
uninterruptible
power supply or a
battery�
Power frequency
(50/60 Hz) magnetic
eld
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical hospital or
home environment�
Note: U
T
is the a�c� mains voltage prior to application of the test level�