60
SimplyGo Mini User Manual
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms
Amateur Radio & ISM
Bands between 150
kHz and 80 MHz
3 Vrms
150 kHz to 80 MHz
6 Vrms
Amateur Radio & ISM
Bands between 150
kHz and 80 MHz
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the device,
including cables, than
the recommended
30 cm (12 inches)
separation distance�
Interference may
occur in the vicinity
of equipment marked
with the following
symbol:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2�7 GHz
Telecommunication
frequencies as
specied in clause
8�10 of IEC 60601-1-2:
2014:
450, 810, 870, 930,
1720, 1845, 1970 and
2450 MHz at 28 V/m
385 MHz at 27 V/m
710, 745, 780, 5240,
5500 and 5785 MHz
at 9 V/m
10 V/m
80 MHz to 2�7 GHz
Telecommunication
frequencies as
specied in clause
8�10 of IEC 60601-1-2:
2014:
450, 810, 870, 930,
1720, 1845, 1970 and
2450 MHz at 28 V/m
385 MHz at 27 V/m
710, 745, 780, 5240,
5500 and 5785 MHz
at 9 V/m

Table of Contents