EasyManuals Logo

Philips Respironics SimplyGo Mini User Manual

Philips Respironics SimplyGo Mini
193 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
5
SimplyGo Mini User Manual
English
Warnings (continued)
Repairs and adjustments must be performed by Philips Respironics
authorized service personnel only� Unauthorized service could cause
injury, invalidate the warranty, or result in costly damage�
Periodically inspect electrical cords, cables, and the power supply for
damage or signs of wear� Discontinue use and replace if damaged�
To avoid electric shock, unplug the device and remove the batteries
before cleaning the device or bag� DO NOT immerse the device in any
uids�
To prevent electrical shock, turn o and disconnect the electrical
supply before attempting to perform any maintenance to the
SimplyGo Mini Concentrator
For proper operation, your device requires unobstructed ventilation�
Always make sure any openings in the case are not obstructed by
items which may impede ventilation� Do not place the device in a
small closed space (such as a closet)� The device should not be used
adjacent to or stacked with other equipment� For more information,
contact your equipment provider
Do not use an extension cord�
Device operation above or outside of the voltage, breath rate,
temperature, humidity and/or altitude values specied may decrease
oxygen concentration levels�
Be aware that the electrical cord and/or tubing could present a
tripping or strangulation hazard�
Use only power cords supplied by Philips Respironics for this device�
Use of power cords not supplied by Philips Respironics may cause
overheating or damage to the device and may result in increased
emissions or decreased immunity of the equipment or system�
The use of accessories, transducers and cables other than those
specied or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper
operation�
Do not operate without the battery installed and working� If primary
power is lost with no battery in place, the device will stop operating
without warning the user� If the device must be operated without the
battery, the user should be aware that there is no backup power and
no warning if primary power is lost�

Table of Contents

Other manuals for Philips Respironics SimplyGo Mini

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Respironics SimplyGo Mini and is the answer not in the manual?

Philips Respironics SimplyGo Mini Specifications

General IconGeneral
Battery Duration (Standard Battery)Up to 4.5 hours
Battery LifeUp to 4.5 hours with standard battery, up to 9 hours with extended battery
Oxygen Concentration87% to 96% at all settings
Pulse Dose Settings1 to 5
Recharge Time (Standard Battery)Approximately 4 hours
Recharge Time (Extended Battery)8 hours
Operating AltitudeUp to 10, 000 feet (3048 meters)
Sound Level43 dBA (at pulse dose setting 2)
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Altitude RangeUp to 10, 000 feet (3048 meters)
Weight5 lbs (2.3 kg) with standard battery
Humidity Range15% to 95% non-condensing

Related product manuals