EasyManua.ls Logo

Planmeca Compact i5 - Page 223

Planmeca Compact i5
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
above the bowl as shown in the figure. Press the outer edge of the
flushing holder cover (1) and remove the cover.
3. Remove the instrument handpieces from the instruments and clean them
according to the manufacturer’s instructions.
4. Place all water consuming instruments (including syringes) into the
openings in the instrument flushing holder.
Optionally, in dental units with balanced instrument arms, you can
remove the hoses of the instruments to be flushed from the instrument
arms / instrument holders before placing the instruments in the flushing
holder.
In dental units with hanging-tube instruments, the instruments to be
flushed must be removed from the instrument holders before placing
them in the flushing holder.
1 Flushing holder cover
Each type of syringe has its dedicated place in the flushing holder. The
picture below shows the syringe openings in the holder, but note that the
picture is an example only and does not represent the actual holder.
28 Flushing and cleaning programs
User's manual Planmeca Compact i 213

Table of Contents

Related product manuals