MacBAT 5 – USER GUIDE. Version 1.0
This paragraph contains of full list of alarms with short description of appearing conditions:
Events typed in bold are the ones which are saved to Alarm LOG memory sector.
CODE
NAME
DESCRIPTION
0
Device startup
Device turned on – initial startup, battery replacement, firmware upgrade or
malfunction.
1
System error
Device has detected the lack of P1 or T sensor
2
Calculation error
PTZ conversion is executed wrong – plausible because of pressure out of range
3
AlgZ range
Gas composition parameters are out of range
4
P1 range
Pressure is out of range – primary sensor
5
P2 range
Pressure is out of range
6
T range
Temperature is out of range
7
Qm range
Measurement flowrate is out of range
8
Tamb error
Ambient temperature sensor failure
9
Tamb range
Ambient temperature is out of range
10
Battery low
Battery level is below 10%
11
Ext. supply off
External power supply source has been disconnected
12
Ext. supply low
Fluctuations of connected power supply source – possible that not compatible unit is
connected – different than INT-S3 or supplied on the desk from USB 5V or another
common power supply
13
Keyboard error
Malfunctioning of the keyboard – device checks the keys status periodically
14
Software update
Device firmware has been updated
15
Data update
Device data has been changed – Modbus maps
16
Data erased
Erasing of the device
17
AlarmLOG full
Alarm LOG is filled up to 97% - requires confirmation of the alarm
18
Case open
Device case is opened
19
Intrussion attempt
5 unsuccessful attempts of login with wrong password. Lockdown of the device COM
ports for 15 minutes
20
Login
User logged in using keyboard