B-12
MacBAT 5 – USER GUIDE. Version 1.0
NAMUR inputs (HF1, HF2) – Terminals 26 (DI6+) to 25 (DI6-), 28 (DI7+) to 27 (DI7-)
Uo=9,6V; Io=33mA; Po=78mW; Li=0; Ci=0
Gas Group IIA: Lo=800mH; Co=100µF
Gas Group IIB: Lo=400mH; Co=13µF
Input SCR ENCODER – Terminals 30 (DI8+), 29 (DI8-) to GND
Uo=9,6V; Io=0,021A; Po=0,48W; Li=0; Ci=0
Gas Group IIA: Lo=2mH; Co=500µF
Gas Group IIB: Lo=1mH; Co=25µF
Sensor Pt1000 – Terminals 32 (I+), 31 (I-), 34 (U+), 33 (U-) to GND
Ui=6,51V; Li=0; Ci=250nF;
External pressure sensor – Terminal 36 (PS1), 38 (PS2), 40 (PS3), 37 (PS4), 39 (PS5) to 35 (GND)
Ui=6,51V; Li=0; Ci=200nF;