Memory for alarms in the device is divided for three sectors:
• Alarm LOG – capacity of the memory – 3000 records. Filling this sector of memory causes generating alarm called Alarm LOG full.
The main effect of full Alarm memory sector is stopping counting to main counters – Vb – volume in base conditions. Configuration
of the device is blocked then. This memory sector requires periodical clearing. Acknowledgement of closed alarms by authorized
personnel – they are acquainted with list of alarms stored in device memory.
Clearing of Alarm memory sector:
o Go to Main menu by pressing Enter
o Navigate to menu Alarms → Configuration
o Press Enter on parameter AlarmLOG
o Set 0 in the value of parameter and confirm changes.
• Process LOG – capacity of the memory – 3000 records. Filling this sector of memory does not generate any alarm. Oldest records
are replaced by the newest one. This memory does not have the influence on measurement and device operation.
• Setup LOG – capacity of the memory – 1000 records. Filling this memory does not generate any alarm. In this memory sector
there are significant interventions saved. Names of the events in this memory sector are not presented.
Clearing of Setup memory sector:
o Go to Main menu by pressing Enter
o Navigate to menu Alarms → Configuration
o Press Enter on parameter SetupLOG
o Set 0 in the value of parameter and confirm changes.