n Maximum cable length: 30 m, screened.
Electrical data
Parameter Value
Cell constant: 0.005 1/cm ... 15 1/cm
Measuring ranges dependent on the sensor type:
Specific conductivity: 0.001 µS/cm ... 200 mS/cm
Specific electrical resistance: 5 Ωcm ... 1000 MΩcm
TOS (total dissolved solids): 0 ... 9999 ppm (mg/l)
SAL (salinity): 0.0 ... 70.0 ‰ (g/kg)
Precision:
Specific conductivity: 1 µS/cm ... 20mS/cm: better 1% of the measured value
±1 µS/cm/±1 digit
Specific electrical resistance: 50 Ωcm ... 10 MΩcm: better 1% of the measured value
±1 digit
Specific electrical resistance: 10 MΩcm ... 100 MΩcm: better 10 MΩcm
Correction variable: Temperature via Pt100/Pt1000, semiconductor temperature sensor
Measuring range: (Pt100/Pt1000: Sensor cable length up
to 10 m)
-20 °C ... +180 °C
Measuring range: (Pt100/Pt1000: Sensor cable length up
to 50 m)
-20 °C ... +120 °C
Measuring range: (Semiconductor temperature sensor) -20 °C ... +125 °C
Precision of the temperature measurement: better 1% of the measured value (maximum 1 °C)