CAL pH
Buffer manufacturer
Calibration process
A1512
Manual
Buffer detection
2 point
Manual
Buffer value 1
Buffer value 2
Buffer temperature
ProMinent
pH 7
pH 4
Buffer temperature 25.0 °C
Fig. 34: Example: Display in[CAL Setup]
Buffer used
Dispose of the used buffer solution.
For more information: refer to the
buffer solution safety data sheet.
Valid calibration values
Valid calibration:
–
Zero point -60mV…+60 mV
–
Slope 55 mV/pH…62 mV/pH
You need two test containers with buffer
solution for calibration. The pH values of
the buffer solutions should be at least 2
pH values apart. Thoroughly rinse the
sensor with water when changing the
buffer solution.
Continuous display ➨
CAL pH
Last calibration
A1016
Zero point
Slope
Calibration process
CAL setup
continue with <CAL>
14:26:07
7,00 pH
2 point
56.64mV/pH
06/04/2013
Fig. 35: pH sensor calibration (CAL)
1.
Then press
2. Rinse the sensor thoroughly with
water before drying with a cloth
(pad dry, don't rub)
3. Immerse the sensor in test con‐
tainer 1 containing buffer solution
(e.g. pH 7). When doing so, move
the sensor gently
Calibration
67