780C Multi-Interface Interoperability Tester User Guide Page 197
September 29, 2014
Revision A1
11.2.3 Procedures for Saving an EDID into the 780C
Use the following procedures to add an EDID to your 780C from a connected HDMI sink device.
1. Make the physical connections between the 780C HDMI or HDBaseT OUT connector and the display
device whose EDID you wish to store.
2. Enable the HDMI output using the procedures at Selecting a Signal Type and Resolution.
3. Touch select EDID Test from the Home menu shown below.
4. Touch select the Read activation button from the EDID Test menu shown below.
The EDID information is presented on multiple pages on the display. An example of a page of the EDID listing
is shown below. You scroll through all the pages using the scroll bar at the right side of the listing.