Remote Control
R&S
®
FPC
327User Manual 1178.4130.02 ─ 07
Suffix:
<n>
.
irrelevant
Return values:
<Peaks> <numeric value>
Example: //Query number of entries in the DTF list
MEAS:MODE DTF
CALC:TRAN:DTF:PEAK:COUN?
Usage: Query only
CALCulate<n>:TRANsform:DTFault:PEAK:DATA<peak>?
This command queries the DTF list.
Prerequisites for this command
●
Select the DTF measurement (MEASurement<n>:MODE).
Suffix:
<n>
.
irrelevant
<peak> Selects an entry in the DTF list.
The range depends on the number of DTF list entries. You can
query the number of entries with CALCulate<n>:TRANsform:
DTFault:PEAK:COUNt?.
Return values:
<Distance>,
<ReturnLoss>
<numeric value>
The first value represents the distance to the fault, the second
value the return loss measured at that distance.
Example: //Query second DTF list entry
MEAS:MODE DTF
CALC:TRAN:DTF:PEAK:DATA2?
Usage: Query only
CALCulate<n>:TRANsform:DTFault:PEAK[:STATe] <State>
This command turns the display of the DTF list on and off.
Prerequisites for this command
●
Select the DTF measurement (MEASurement<n>:MODE).
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Remote Commands of the VNA Application