Remote Control
R&S
®
FPC
326User Manual 1178.4130.02 ─ 07
CALCulate<n>:DTF:DISTance:STARt <StartDistance>
This command defines the starting point (distance) of the DTF measurement.
Prerequisites for this command
Select the DTF measurement (MEASurement<n>:MODE).
Suffix:
<n>
.
irrelevant
Parameters:
<StartDistance> <numeric value>
The unit depends on your selection (UNIT<n>:LENGth).
*RST: 0 m
Default unit: m or ft
Example: //Define start distance for DTF measurement
UNIT:LENG MET
MEAS:MODE DTF
CALC:DTF:DIST:STAR 10
Manual operation: See "Start Distance" on page 398
CALCulate<n>:DTF:DISTance:STOP <StopDistance>
This command defines the end point (distance) of the DTF measurement.
Prerequisites for this command
Select the DTF measurement (MEASurement<n>:MODE).
Suffix:
<n>
.
irrelevant
Parameters:
<StopDistance> <numeric value>
The unit depends on your selection (UNIT<n>:LENGth).
*RST: 50 m
Default unit: m or ft
Example: //Define stop distance for DTF measurement
UNIT:LENG MET
MEAS:MODE DTF
CALC:DTF:DIST:STOP 100
Manual operation: See "Stop Distance" on page 398
CALCulate<n>:TRANsform:DTFault:PEAK:COUNt? <State>
This command queries the number of entries in the DTF list.
Prerequisites for this command
Select the DTF measurement (MEASurement<n>:MODE).
Remote Commands of the VNA Application