EasyManua.ls Logo

R&S FPC Series - Page 357

R&S FPC Series
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FPC
354User Manual 1178.4130.02 ─ 07
The ENABle part allows true conditions in the EVENt part of the status register to be
reported in the summary bit. If a bit is 1 in the enable register and its associated event
bit transitions to true, a positive transition will occur in the summary bit reported to the
next higher level.
Parameters:
<SumBit> <numeric value> (integer)
Range: 0 to 65535
STATus:QUEStionable:FREQuency:NTRansition <SumBit>
STATus:QUEStionable:LIMit:NTRansition <SumBit>
STATus:QUEStionable:POWer:NTRansition <SumBit>
STATus:QUEStionable:NTRansition <SumBit>
This command controls the Negative TRansition part of the QUEStionable register.
Setting a bit causes a 1 to 0 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Parameters:
<SumBit> <numeric value> (integer)
Range: 0 to 65535
STATus:QUEStionable:FREQuency:PTRansition <SumBit>
STATus:QUEStionable:LIMit:PTRansition <SumBit>
STATus:QUEStionable:POWer:PTRansition <SumBit>
STATus:QUEStionable:PTRansition <SumBit>
This command control the Positive TRansition part of the QUEStionable register.
Setting a bit causes a 0 to 1 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Parameters:
<SumBit> <numeric value> (integer)
Range: 0 to 65535
Status Reporting System