Attention:
No button may be activated on the unit during the update process. Pressing a button can disrupt the
updating program or even cancel it. The existing operating software may then be defective. The unit no
longer responds. The batteries/rechargeable batteries must be removed from the battery compartment
for a RESET or the unit must be sent back to the manufacturer for repair.
If the update should fail for other reasons, the update program is
interrupted and a corresponding error message is displayed on the
screen.
This must be read and confirmed by pressing the <OK> button.
The unit then goes back to the setup menu screen. An update has
not been carried out, however. If you cannot resolve the error
yourself, the details of the error message should be notated and
the manufacturer contacted.
Return to the main menu with <ESC>.
3.8.2 Updating with the "Update Program" from the
REA CD-ROM
To carry out an update, the PC program must be loaded onto a PC and started. The program required
for this is called UpdateServer.Exe.
It can be downloaded separately from the enclosed CD-ROM to a PC with the Windows operating
system or started from the program REA TransWin32 from the menu bar (see operating instructions of
REA TransWin32 for installation instructions).
To use this software, network communication must be established between REA ScanCheck 3n and
the network interface on the PC. The supplied network cable and the PoE power supply are required
for the connection (see Chapter 2).
Once the program has started, the following screen
appears on the desktop:
The new update file must first be found with the
browse function on the PC and highlighted.
Press the <...> button after the "Update File" field.
The PC file explorer opens. Find the corresponding
directory and select the desired update file
"updatesc3.tgz" by clicking on the right mouse
button.
Confirm by pressing the <Open> button.