Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.140 E-6
The Define Segments dialog contains a table to edit the individual segments of the sweep range.
Sweep segments must not overlap, however, two adjacent segments can have one common point; see
Columns in the Define Limit Line table.
Below the table, three groups of controls provide additional settings. Refer to the following sections
below:
• Inserting and Deleting Segments
• Individual Segment Settings
• Point List, Segment Import and Export
Columns in the Define Limit Line table
The table contains an automatically assigned current number for each segment plus the following
editable or non-editable columns:
• On provides check boxes to activate or deactivate each individual segment. Sweep points
belonging to inactive segments only are not measured and not listed in the Point List.
• Start is the stimulus (x-axis) value of the first point of the segment. If the segment contains
more than one Point, then Start must be smaller than the Stop value. If a Start value equal to or
larger than the current Stop value is set, Stop is adjusted to the new Start value plus 1 Hz.
• Stop is the stimulus (x-axis) value of the last point of the segment. If the segment contains more
than one Point, then Stop must be larger or equal than the Start value. If a Stop value equal to
or smaller than the current Start value is set, Start is adjusted to the new Stop value minus 1
Hz.