R&S
®
ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.179 E-6
4. Receiver power calibration for the IF signal. No external device is needed. The receiver
(port 2) is calibrated at the IF frequency using the source signal from port 1 (red) calibrated in
the first step.
5. Source power calibration for the LO signal, provided by an external generator. A power
sensor is connected to the generator's RF output (red signal). The analyzer controls the
generator, its RF power (red) is calibrated at the selected IF frequency. No RF connection to or
from the network analyzer is needed.