EasyManua.ls Logo

Rohde & Schwarz R&S ZVA 8 - Page 7

Rohde & Schwarz R&S ZVA 8
813 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1145.1084.02 RE.7 E-6
Automatic power reduction for Calibration Unit in the initial tab of the System Config dialog
GUI improvements
Paste marker list for easier data entry in numeric entry bars and dialogs.
System files (*.zvx, *.s<n>p, *.csv, *.ck, *.calkit, *.seglist, *.mth, *.limit, *.colorscheme) can be
loaded by drag and drop or double-click in the Windows Explorer.
New remote control features
The new features are also available via remote control; the SCPI commands are reported in the
relevant reference sections. Besides the following remote-control features have been added:
User defined color scheme settings (DISPlay:CMAP...).
Cal standard data can be loaded from a Touchstone file (MMEMory:LOAD:CKIT:SDATa...)
New command SYSTem:ERRor:ALL reads complete error queue.
New command FORMat:BORDer controls whether binary data is transferred in normal or
swapped byte order.
Fixed Issues in Firmware V1.62 (Compared to V1.61)
Improved stability of the TNA calibration.
Corrected TOM, TNA, TRM, TRL error term conversion improves the accuracy of the 7-term
calibration types.
Fixed Issues in Firmware V1.62 (Compared to V1.61)
Time domain transform of impedances and admittances corrected.
New Features in Firmware 1.60 (Compared to V1.51)
Support for new option: Time Domain (option R&S ZVAB-K2)
Handling of software options: Option key entry in System Config. menu, option Info
Double-click magnifies diagram areas
Fixed Issues
The limitation concerning the port groups reported for FW V1.50 no longer applies. Port groups can be
measured and used simultaneously.
Fixed Problems in Firmware V1.51 (Compared to V1.50)
Mixed mode parameters corrected.
Problems with VX11 GPIB bus connection solved.

Table of Contents