R&S
®
ZVA/ZVB/ZVT SOURce<Ch>:POWer:CORRection...
1145.1084.12 6.277 E-6
SOURce<Ch>:POWer<Pt>:CORRection:MIXer:LO[:ACQuire]
Starts the LO source calibration (3
rd
power calibration step for mixer measurements), stores and applies
the calibration data. The external power meter is selected via
SOURce<Ch>:POWer<Pt>:CORRection:PMETer:ID.
The LO source calibration is independent of the RF source and IF receiver calibration; see
description of the mixer calibration procedure.
<Ch>
Calibrated channel number
<pt>
Calibrated port number. This suffix is ignored; the port is selected via
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOINternal or
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOEXternal.
SCPI,
Command
Types
Device-specific, no query
Example:
See SOURce<Ch>:POWer<Pt>:CORRection:MIXer:RF[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:MIXer:RF[:ACQuire]
Starts the RF source calibration (1
st
power calibration step for mixer measurements), stores and applies
the calibration data. The external power meter used is selected via
SOURce<Ch>:POWer<Pt>:CORRection:PMETer:ID.
The command initiates two calibration steps, the first at the RF frequency range, the second at
the IF frequency (for a subsequent receiver power calibration,
[SENSe<Ch>:]CORRection:POWer<port_no>:MIXer:IF:ACQuire). See description of the
mixer calibration procedure.
<Ch>
Calibrated channel number
<pt>
Calibrated port number. This suffix is ignored because the RF signal is
always assigned to port 1.
SCPI,
Command
Types
Device-specific, no query