SOURce<Ch>:POWer<Pt>:CORRection:PMETer:ID <pmeter_no>
Selects an external power meter for the source power calibration.
The command cannot be used unless a power meter is connected via GPIB bus, USB or LAN
interface and configured in the External Power Meters dialog.
<Ch>
Calibrated channel number
<Pt>
Calibrated port number. This parameter is ignored; the power meter
selection is valid for all sources.
<pmeter_no>
Number of external power meter. The parameters UP, DOWN, MIN,
MAX are not available for this command.
Range [def. unit] 1 to number of external generators [–]
*RST value The power meter selection is not changed by a reset of the analyzer.
SCPI,
Command
Types
Device-specific, command or query.
Example:
See SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire] and
SOURce<Ch>:POWer<Pt>:CORRection:MIXer:RF[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection[:STATe] <Boolean>
Enables or disables the source power calibration for channel <Ch> and for port number <Pt>. The
command is disabled unless a source power calibration for the analyzer port has been performed
(SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]). To enable or disable a source power
calibration for an external generator use
SOURce<Ch>:POWer<Pt>:CORRection:GENerator[:STATe].
<Ch>
Calibrated channel number
<Pt>
Calibrated port number
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number <Pt>.
*RST value OFF [–]
SCPI, Command
Types
Device-specific, command or query
Example:
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].